Item: 50104242
50104242 XS PC 80 PRO Stirrer - Bench multiparameter - Cell VPT1 - Electrode Standard S7
Item: 50104242
50104242 XS PC 80 PRO Stirrer - Bench multiparameter - Cell VPT1 - Electrode Standard S7
Technical specifications | |
---|---|
ISE accuracy | 0.5 % f. s. for monovalent ions - 1 % for divalent ions |
Accuracy (pH) | ±0,002 |
Accuracy (Salinity) | ±1 % f.s. |
Accuracy (Temperature) | ±0,2 |
Magnetic stirrer | Yes, independently controlled |
Temperature calibration | Yes |
AQA calibration | Yes |
Temperature coefficent (max) | 0,00 ... 10,00 |
DHS compatibility | Yes |
Temperature compensation | 0,0 ... +100,0 |
Ultrapure water compensation | Yes for values below 10 µS |
PC Communication | USB type B |
Printer connection | |
Stirring speed control | 15 levels |
Cell constant | 2 rings: 0.1 / 1 / 10 | 4 rings: 0.55 |
Criteria for stability of the measure | Med - High - Tit | Secondi (ISE) |
Dimension | |
TDS factor | 0,40 ... 1,00 |
Warranty | |
AQA | Yes |
Inputs | 2 x BNC - 2 x RCA - Multipin - 4 mm banana |
Selectable languages | Eng - Ita - Deu - Esp - Fra - Por |
Memory | 8000 |
Parameters | pH - mV - ORP - ISE - Cond - TDS - Sal - Res - T |
Password | 6 to 10 characters with a capital letter, a number and a symbol |
Weight | 1,255 |
Ion measuring range | 0,001 ... 19999 |
mV ORP measuring range | -2000 ... +2000 |
PH calibration points | 1 ... 5 automatic values - 1 ... 5 user values |
pH measuring range | -2,000 ... +20,000 |
Resistivity measuring range | 1,00 Ω ... 10,00 MΩ |
Salinity measuring range | 0,01 ppm … 100 ppt |
Cond resolution | 0,01 / 0,1 / 1 (scala automatica) |
ORP resolution | 0.1 / 1 (automatic) |
pH resolution | 0,1 / 0,01 / 0,001 |
Temperature resolution | 0,1 |
Buzzer | Yes |
Timer calibration due with alarm | Si |
Cell type | 2 rings - 4 rings |
ION measuring unit | mg/l - g/l - mol/l |
Stirring speed | 0 ... 3000 |
Max stirring quantity | 2 |
Power supply | 100 - 240 |
Analog view | Yes |
Accuracy (TDS) | ±1 % f.s. |
pH alarm range | Min - Max |
Recognised pH buffers | USA - NIST - DIN 19266 - Mettler Toledo Eu - Merck |
Swabs recognised Cond | 84 µS - 147 µS - 1413 µS - 12,88 mS - 111,8 mS |
Working temperature | 0 ... +45 |
Reference temperature | +15 ... +30 |
Accuracy (Cond) | ±0,5 % f.s. |
Display | 4.3" high-definition, backlit colour TFT |
Calibration data indication | Yes with reports |
Buffer used indication | Yes with icon |
Electrode status indication | Yes with graph and icon |
Calibration points Cond | 1 ... 4 automatic values - 1 user value |
ION calibration points | 2 ... 5 |
ORP calibration points | 1 automatic value - 1 user value |
Cond measuring range | 0,00 μS ... 1000 mS |
TDS measuring range | 0,1 mg/l ... 500 g/l |
temperature measurement range | -30,0 ... +130,0 |
Calibration Report | Yes, printable |
Customised Print Reports | 4 customisable print templates |
Access Profiles | 3 livelli per 15 utenti; livello Guest libero |
ORP alarm thresholds | Min - Max |
Simultaneous Display pH_Cond | Yes |
Entering identifiers | Yes via external keyboard |
Audit Trail | Yes via PC software |
Multiview | Yes, up to 6 parameters simultaneously |
Data storage modes | Manual - Automatic - Auto with Min/Max threshold - Auto with delta value |
TDS unit of measurement | mg/l - g/l | ppm - ppt |
Unit of measurement Salinity | ppm - ppt | mg/l - g/l |